Bovie® IDS-310 300W Electrosurgical Generator

Part # IDS-310
IDS-310-wNewBovieLogo-Front-View
IDS-310-wNewBovieLogo-Front-View
QTY
  • Technical Specifications
    MONOPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
    Cut I 300 1000V 1.7 ± 20% 490 kHz ± 4.9 kHz
    Cut II 300 750V 1.7 ± 20% 490 kHz ± 4.9 kHz
    Blend (1) 200 1320V 1.8 ± 20% 490 kHz ± 4.9 kHz
    Blend (2) 200 1475V 2.0 ± 20% 490 kHz ± 4.9 kHz
    Blend (3) 200 1650V 2.2 ± 20% 490 kHz ± 4.9 kHz
    Blend (4) 200 1870V 2.4 ± 20% 490 kHz ± 4.9 kHz
    Pinpoint Coagulation 120 1800V 3.1 ± 20% 490 kHz ± 4.9 kHz
    Spray Coagulation 120 4000V 6.0 ± 20% 350 to 450 kHz
    Gentle Coagulation 120 450V 1.6 ± 20% 490 kHz ± 4.9 kHz
    BIPOLAR WATTS VPEAK MAX VOLTAGE CREST FACTOR (TYPICAL) FREQUENCY (KHZ)
    Macro Bipolar 150 600V 1.5 ± 20% 490 kHz ± 4.9 kHz
    Micro Bipolar 80 450V 1.5 ± 20% 490 kHz ± 4.9 kHz
    Standard Bipolar 50 300V 1.5 ± 20% 490 kHz ± 4.9 kHz
    Bovie Bipolar 225 250V 1.5 ± 20% 490 kHz ± 4.9 kHz
    Auto Bipolar 80     490 kHz ± 4.9 kHz

    Technical Specifications:
    • Lifespan: Reusable
    • Sterilization Status: Non-sterile

    Electrical Specifications:
    • Cooling System: Passive Cooling
    • Technology: Isolated RF Output
    • Audible/Visible Output Meter: Yes
    • Pre-set Feature: Yes
    • Monopolar HP Ports: 2 Port(s)
    • Connector Ports/Power Switch: Black
    • Contact Monitoring Systems: Bovie NEM™
    • Contact Monitoring Systems Range/Trip Point: 10 ohms or less over entire operating range (10C° to 40°C)
    • Dual Activation Tone Alert: Yes

    Safety Specifications:
    • Latex-Free: Yes
  • Included Items
  • Resources
  • Features

    ENERGY SETTINGS

     
     

    CUT

    Cut I

    A general cutting with a standard RF waveform used in open and endoscopic surgical procedure, Cut I is for use in a wide range of tissue with minimal hemostasis.

    Cut II

    Tailored specifically for Laparoscopic procedures, Cut II offers a precisely managed cutting energy that limits output for the minimizing of the potential risk of capacitive coupling and alternate site burns due to weakened insulation.

    Blend (Level 1, 2, 3 & 4)

    Blended cutting offers varying levels of hemostasis during a cut providing increased control of bleeding at the surgical application site.

    • Level 1: 75% Duty Cycle
    • Level 2: 62.5% Duty Cycle
    • Level 3: 50% Duty Cycle
    • Level 4: 37.5% Duty Cycle
     
     
    Cut
     

    COAGULATION

    Pinpoint

    A general coagulation mode delivering rapid desiccation effects with minimal tissue damage.

    Spray

    Recommended for hemostasis in highly vascular or large surface areas and for non-contact tissue coagulation to control superficial bleeding. Spray mode permits dual, simultaneous activation of two monopolar accessories with audible dual activation tone alert.

    Gentle

    Gentle mode offers more concentrated coagulation than in other modes. For instance, when coagulation is necessary in short amounts of time, reduced electrode carbonization is provided.

     
     
    Coag
  • Overview

    BOVIE BIPOLAR (BBP)

    Bovie Bipolar is a pulsing coagulation for quick and powerful coagulation of tissue. The pulsing performance can be tailored specifically to achieve the optimal surgical effect.

    AUTO BIPOLAR INTUITIVE TECHNOLOGY

    Auto Bipolar offers immediate and simplified tissue coagulation with minimal tissue damage. This feature automatically activates, coagulates and deactivates for precise tissue coagulation and eliminates the need for a footswitch. This feature is compatible with Bovie’s Micro, Macro and Standard Bipolar modes and can be set to surgical preferences for critical procedures.

    SPECIALIZED LAPAROSCOPIC MODE

    The IDS-310 recalls the last used power and mode. The unit’s user preset can recall up to 10 presets and up to 6 bipolar output preferences.

    ADJUSTABLE POWER DURING ACTIVATION

    During all activations, the power can be adjusted up or down in variable steps, giving surgeons more control at the surgical site.

    PATIENT SAFETY FEATURES

    BovieNEM™ – This return electrode contact quality monitoring system is designed to minimize patient burns at the return electrode site. Bovie FDFS™ – The tissue sensing technology measure tissue 5 million times a second and adjusts to varying impedances 5000 times a second for accurate energy delivery. A visual Ammeter indicates real time current (mA) flowing through the patient.